GUI reference
R&S
®
ZNA
616User Manual 1178.6462.02 ─ 20
available as a signal source in the test setup. The analyzer provides two alternative,
independent ways of selecting a generator as a signal source:
Check "Source Gen" in the "Port Settings" dialog, especially if the generator is not
assigned to a particular measurement result or drive port.
Select the generator as a source for a particular measurement result or for a power
calibration, see e.g. Chapter 5.2.3.2, "More Wave Quantities dialog", on page 316.
Remote command:
SOURce<Ch>:POWer<PhyPt>:PERManent[:STATe]
SOURce<Ch>:RLO:PERMenable
SOURce<Ch>:POWer<PhyPt>:GENerator<Gen>:PERManent[:STATe]
Freq. Conversion
Opens a dialog to define a port-specific stimulus frequency range (for frequency
sweeps) or CW frequency (for power, time and CW mode sweeps); see "Modify Fre-
quency Conversion dialog" on page 617. In the default configuration, the channel
base frequency f
b
is used.
This cell is only enabled, if the analyzer is equipped with option R&S ZNA-K4. For fre-
quency converter ports (with option R&S ZNA-K8), the formulas for the RF IN and "LO
IN" components are derived from the converter settings and hence are read-only.
The result is displayed in the "Frequency Result" column. A red "Freq. Conversion" cell
indicates that certain hardware limits are exceeded.
Remote command:
SOURce<Ch>:FREQuency<PhyPt>:CONVersion:ARBitrary:IFRequency
SOURce<Ch>:RLO:FREQuency
Receiver Freq.
This cell is only enabled, if the analyzer is equipped with option R&S ZNA-K4. It
defines the frequencies to which the receivers are set while this port is the source port.
"Src. Freq."
(default)
Both receivers use the source formula (see "Freq. Conversion"
on page 616).
"a,b"
Both receivers use the receiver formula (see "Freq. Conversion"
on page 616).
b
The a wave receiver uses the source formula, the b wave receiver the
uses the receiver formula
Remote command:
[SENSe<Ch>:]FREQuency<Pt>:CONVersion:AWReceiver[:STATe]
Receiver Freq. Conversion
This cell is only enabled, if Receiver Freq. is not set to "Src Freq.".
It opens a dialog to define a port-specific receiver frequency range (for frequency
sweeps) or CW frequency (for power, time and CW mode sweeps); see "Modify Fre-
quency Conversion dialog" on page 617. In the default configuration, the source fre-
quency range is used.
For frequency converter ports (with option R&S ZNA-K8), the formulas for the RF IN
and "LO IN" components are derived from the converter settings and hence are read-
only.
Channel Config softtool