Command reference
R&S
®
ZNA
1431User Manual 1178.6462.02 ─ 20
SOURce<Ch>:POWer<Pt>:ALC:COUPle <Boolean>
Enforces identical ALC settings for the individual ports in channel <Ch>.
Suffix:
<Ch>
.
Channel number
<Pt> Port number. This suffix is ignored; the setting applies to all
ports.
Parameters:
<Boolean> ON (1)
Enforces identical ALC settings
OFF (0)
Enables port-specific ALC settings
Manual operation: See "Couple Port Settings" on page 480
SOURce<Ch>:POWer<Pt>:ALC:OPTimize <Alc optimize>
Suffix:
<Ch>
.
<Pt>
Parameters:
<Alc optimize> AUTO | ON | OFF
Manual operation: See "ALC Optimize" on page 481
SOURce<Ch>:POWer<Pt>:ALC:PIParameter <Mode>
Enables automatic or manual setting of the ALC PI controller parameters at source port
<Pt> in channel <Ch>.
See "Control loop parameters" on page 92.
Suffix:
<Ch>
.
Channel number
<Pt> Port number
Parameters:
<Mode> AUTO | MANual
AUTO
The control loop parameters are calculated by the analyzer firm-
ware.
MANual
The parameters can be set using SOURce<Ch>:POWer<Pt>:
ALC:PIParameter:GAIN and SOURce<Ch>:POWer<Pt>:
ALC:PIParameter:ITIMe, respectively.
*RST: AUTO
Manual operation: See "Control Loop Parameters" on page 482