Command reference
R&S
®
ZNA
1438User Manual 1178.6462.02 ─ 20
Manual operation: See "Number of Points" on page 549
SOURce<Ch>:POWer<PhyPt>:CORRection:SLEVeling:FREQuency:STARt
<Leveling Tool Start Frequency>
SOURce:POWer<PhyPt>:CORRection:SLEVeling:FREQuency:STOP <FreqValue>
Defines the start and stop frequency of the leveling grid for physical port <PhyPt>.
By default the vector network analyzer uses the standard frequency range of the
selected "Meas Port".
After defining the leveling grid (using
SOURce<Ch>:POWer<PhyPt>:CORRection:SLEVeling:FREQuency and
SOURce<Ch>:POWer<PhyPt>:CORRection:SLEVeling:POWer commands), you
can start the leveling data acquisition using SOURce:POWer<PhyPt>:CORRection:
SLEVeling[:ACQuire].
Suffix:
<PhyPt>
.
Physical port number
Parameters:
<FreqValue> Start/stop value
Default unit: Hz
Example: See SOURce:POWer<PhyPt>:CORRection:SLEVeling[:
ACQuire]
Options: R&S ZNA-K8
Manual operation: See "Start Frequency/Stop Frequeny" on page 549
SOURce:POWer<PhyPt>:CORRection:SLEVeling:POWer:ALC <Compatible>
If checked, the power step size of the leveling grid for physical port <PhyPt> is fixed to
1 dB. The step size defined using SOURce:POWer<PhyPt>:CORRection:
SLEVeling:POWer:STEP is ignored.
After defining the leveling grid (using
SOURce<Ch>:POWer<PhyPt>:CORRection:SLEVeling:FREQuency and
SOURce<Ch>:POWer<PhyPt>:CORRection:SLEVeling:POWer commands), you
can start the leveling data acquisition using SOURce:POWer<PhyPt>:CORRection:
SLEVeling[:ACQuire].
Suffix:
<PhyPt>
.
Physical port number
Parameters:
<Compatible>
Example: See SOURce:POWer<PhyPt>:CORRection:SLEVeling[:
ACQuire]
Options: R&S ZNA-K8
Manual operation: See "ALC Step Size" on page 550
SCPI command reference