Command reference
R&S
®
ZNA
1453User Manual 1178.6462.02 ─ 20
Suffix:
<Ch>
.
Calibrated channel number
<PhyPt> Calibrated port number. This suffix is ignored; the generator is
selected via <Gen>.
<Gen> Generator number
Parameters:
<Boolean> Enables (ON) or disables (OFF) the source power calibration for
port number <PhyPt>.
*RST: OFF
Example: See SOURce<Ch>:POWer<PhyPt>:CORRection[:ACQuire]
Manual operation: See "Port Overview" on page 582
SOURce<Ch>:POWer:CORRection:IMODulation:METHod
<FastOrNot>
Enables or disables the fast interpolation method for intermodulation calibration.
Suffix:
<Ch>
.
Channel number
Parameters:
<FastOrNot> ON (1)
Fast method enabled
OFF (0)
Fast method disabled
*RST: ON (1)
Options: R&S ZNA-K4
Manual operation: See "IMD Cal Method>Fast Cal Interp. Meas" on page 537
SOURce<Ch>:POWer<PhyPt>:CORRection:IMODulation:LTONe[:ACQuire]
Starts the source calibration for the lower tone (1
st
power calibration step for intermo-
dulation measurements), stores and applies the calibration data. The external power
meter used is selected via SOURce<Ch>:POWer<PhyPt>:CORRection:PMETer:ID.
Suffix:
<Ch>
.
Calibrated channel number
<PhyPt> Port number This suffix is ignored; the analyzer calibrates the
port that is selected as a source port for the intermodulation
measurement
([SENSe<Ch>:]FREQuency:IMODulation:LTONe).
SCPI command reference