Command reference
R&S
®
ZNA
1452User Manual 1178.6462.02 ─ 20
Example:
*RST; :SWE:POIN 10
Reset the instrument and reduce the number of sweep points to
10.
SOUR:POW:CORR:COLL:DEF ON
Enable a dummy source power calibration.
SOUR:POW:CORR:ACQ PORT,1
Perform a dummy source power calibration using port 1 as a
source port.
SOUR:POW:CORR:DATA? 'A1'
Query the correction values. The analyzer returns 10 comma-
separated real numbers.
MMEM:STOR:CORR 1,'Calgroup_Ch1.cal'
Store the default source power calibration data to the calibration
pool in the default directory C:\Rohde&Schwarz\NWA\Calibra-
tion\Data.
<Replace the generated file Calgroup_Ch1.cal with your exter-
nal power calibration data.>
MMEM:LOAD:CORR 1,'Calgroup_Ch1.cal'
Load the external data and apply them to channel 1 and port 1.
DISP:MENU:KEY:EXEC 'Cal Manager'
Open the Calibration Manager dialog to check the result of the
previous actions.
SOURce<Ch>:POWer<PhyPt>:CORRection:FAST <arg0>
Enables or disables a fast source power calibration, where the external power meter is
used for the first calibration sweep only.
It is recommended to use the more general command SOURce<Ch>:POWer<PhyPt>:
CORRection:COLLect:METHod.
Suffix:
<Ch>
.
Calibrated channel number. This parameter is ignored; the fast
calibration mode applies to all channels.
<PhyPt> Calibrated port number. This parameter is ignored; the fast cali-
bration mode is either enabled or disabled for all sources.
Parameters:
<arg0> Enables (ON) or disables (OFF) fast source power calibration.
*RST: OFF
SOURce<Ch>:POWer<PhyPt>:CORRection:GENerator<Gen>[:STATe] <Boolean>
Enables or disables the source power calibration for channel <Ch> and for an external
generator number <Gen>. The command is disabled unless a source power calibration
for the external generator has been performed (SOURce<Ch>:POWer<PhyPt>:
CORRection[:ACQuire]). To enable or disable a source power calibration for an
analyzer port use SOURce<Ch>:POWer<PhyPt>:CORRection:STATe.