Command reference
R&S
®
ZNA
1457User Manual 1178.6462.02 ─ 20
PPOWer
Predefined port power value, see SOURce:POWer:
CORRection:PPOWer
*RST: CPOWer
Example: See SOURce<Ch>:POWer<PhyPt>:CORRection[:ACQuire]
.
Manual operation: See "Reference Receiver Cal Power" on page 576
SOURce<Ch>:POWer<PhyPt>:CORRection:STATe
<Boolean>
Enables or disables the source power calibration for channel <Ch> and for port number
<PhyPt>. The setting command is disabled unless a source power calibration for the
analyzer port has been performed (SOURce<Ch>:POWer<PhyPt>:CORRection[:
ACQuire]). The query always returns a result.
To enable or disable a source power calibration for an external generator use
SOURce<Ch>:POWer<PhyPt>:CORRection:GENerator<Gen>[:STATe].
Suffix:
<Ch>
.
Calibrated channel number
<PhyPt> Calibrated port number
Parameters:
<Boolean> Enables (ON) or disables (OFF) the source power calibration for
port number <PhyPt>.
*RST: OFF
Example: See SOURce<Ch>:POWer<PhyPt>:CORRection[:ACQuire]
.
Manual operation: See "Port Overview" on page 582
SOURce<Ch>:POWer<PhyPt>:CORRection:TCOefficient[:STATe] <Boolean>
Enables or disables the use of two-port transmission coefficients.
Suffix:
<Ch>
.
Calibrated channel number. This parameter is ignored; the trans-
mission coefficient settings are valid for all channels.
<PhyPt> Calibrated port number. This parameter is ignored; the transmis-
sion coefficient settings are valid for all sources.
Parameters:
<Boolean> Enables (ON) or disables (OFF) the two-port transmission coeffi-
cients.
*RST: n/a
Example: See SOURce<Ch>:POWer<PhyPt>:CORRection:
TCOefficient:CALibration
SCPI command reference