Command reference
R&S
®
ZNA
1527User Manual 1178.6462.02 ─ 20
<inp> 1, 2
Distinguishes between the two trigger events that can be logi-
cally combined to trigger measurement sequence <Sequence>
(see TRIGger<Ch>[:SEQuence]:MULTiple:LOGic
on page 1525).
Parameters:
<Sequence> SWEep | SEGMent | POINt | PPOint
Triggered measurement sequence
PPOint denotes "partial measurement"; see TRIGger<Ch>[:
SEQuence]:LINK.
<TrigSource> IMMediate | EXTA | EXTB | EXTC | EXTD | MANual | UPORt2
Trigger source for the triggered measurement sequence
IMMediate
Free run measurement (untriggered)
EXTA
Trigger event at Trigger In A or Pin 2 of User Port
EXTB | EXTC | EXTD
Trigger event at Trigger In B, ..., Trigger In D
(requires Trigger board hardware option R&S ZNA-B91)
UPORt2
Trigger event at pin 25 of User Port
MANual
Trigger event generated by pressing the "Manual Trigger" soft-
key
Example:
See TRIGger<Ch>[:SEQuence]:MULTiple:LOGic
on page 1525
Manual operation: See "... /Logic/ ..." on page 509
TRIGger<Ch>[:SEQuence]:SLOPe <Slope>
Qualifies whether the trigger event occurs on the rising or on the falling edge or on the
beginning of the high / low level periods of the external TTL trigger signal.
Suffix:
<Ch>
.
Channel number
Parameters:
<Slope> POSitive | NEGative | HIGH | LOW
Trigger slope for the triggered measurement sequence:
POSitive | NEGative - rising or falling edge
HIGH | LOW - high or low level
Example:
TRIG:SOUR EXT
Activate external signal as trigger source.
TRIG:SLOP NEG
Trigger on the negative edge of the (external TTL) trigger signal.