Command reference
R&S
®
ZNA
1526User Manual 1178.6462.02 ─ 20
OR
logical OR of trigger events 1 and 2
Example:
TRIGger<Ch>:SOURce MULTiple
Activate multiple trigger mode.
:TRIG:MULTiple:SOUR1 SWE, EXTA; SOUR2 SWE, EXTB
Define Trigger In A and B as trigger sources for sweeps.
TRIG:MULT:SLOP1 SWE, POS; SLOP2 POIN, POS
Trigger on the rising edges of Trigger In A and B.
TRIG<Ch>:MULT:LOG SWE, AND
Trigger a new sweep if both trigger events have occurred.
TRIG:MULT:HOLD POIN, 1ms
Define a trigger delay of 1 ms before each sweep start.
Manual operation: See "... /Logic/ ..." on page 509
TRIGger<Ch>[:SEQuence]:MULTiple:SLOPe<Num> <MeasSequence>[, <Slope>]
Qualifies whether the multiple trigger events occur on the rising or on the falling edge
or on the beginning of the high / low level periods of the external TTL trigger signal
<Num>.
Suffix:
<Ch>
.
Channel number
<Num> Number of trigger signal (1 or 2)
Parameters:
<MeasSequence> SWEep | SEGMent | POINt | PPOint
Triggered measurement sequence, PPOint denotes "partial
measurement"; see TRIGger<Ch>[:SEQuence]:LINK.
<Slope> POSitive | NEGative | HIGH | LOW
Trigger slope for the triggered measurement sequence:
POSitive | NEGative - rising or falling edge
HIGH | LOW - high or low level
Example: See TRIGger<Ch>[:SEQuence]:MULTiple:SOURce<inp>
Manual operation: See "... /Logic/ ..." on page 509
TRIGger<Ch>[:SEQuence]:MULTiple:SOURce<inp> <Sequence>[, <TrigSource>]
Selects the source of the trigger events the analyzer uses to start a measurement
sequence in multiple trigger mode (TRIGger<Ch>[:SEQuence]:SOURce MULTi
ple).