Command reference
R&S
®
ZNA
1531User Manual 1178.6462.02 ─ 20
BEFore
is appropriate if the device needs to be triggered before the data
is acquired (typical example: a power meter which is used for
power calibration)
AFTer
is appropriate if the device needs to be triggered just after data
acquisition (typical example: an external generator which must
be reconfigured to get ready for the next measurement)
*RST: AFTer
Example: See TRIGger:CHANnel<Ch>:AUXiliary<n>:STYPe
on page 1531
Options: R&S ZNA-B91
Manual operation: See "Position" on page 510
TRIGger:CHANnel<Ch>:AUXiliary<n>:STYPe
<Type>
Defines how the selected sweep sequence is indicated at the related Trigger Out con-
nector.
Use TRIGger:CHANnel<Ch>:AUXiliary<n>:INTerval to select the sweep
sequence.
Suffix:
<Ch>
.
Channel number
<n> 1 = Trigger Out A, ..., 4 = Trigger Out D
Parameters:
<Type> POSitive | NEGative | AHIGh | ALOW
POSitive | NEGative
pulses with positive or negative polarity
AHIGh | ALOW
High (low) voltage level in active state and low (high) level other-
wise
Note that pulse generator pulses cannot be indicated using high
and low voltage levels.
Example: See TRIGger:CHANnel<Ch>:AUXiliary<n>:INTerval
on page 1530.
Options: R&S ZNA-B91
Manual operation: See "Signal Type" on page 509
7.3.19.3 Other commands
TRIGger:STATus:READy? [<TrigSource>]
Queries the current "Ready for Trigger" state