[SENSe<Ch>:]FREQuency:OFFSet:PWAVes:STATe <Boolean>
Enables/disables the calculation of primed wave quantities for two-tone group delay
measurements.
Suffix:
<Ch>
.
Channel number
Parameters:
<Boolean> ON | OFF | 0 | 1
*RST: 1
Options: R&S ZNA-K9
Manual operation: See "Use Primed Waves" on page 635
[SENSe<Ch>:]FREQuency<Pt>:OFFSet:PWAVes
<FreqOffset>
Defines the constant receiver frequency offset for primed wave quantities a' and b' rela-
tive to the common receiver frequency.
Suffix:
<Ch>
.
Channel number
<Pt> Physical port number
Parameters:
<FreqOffset> Frequency offset.
Range: 0 Hz to 25 MHz
Increment: 1 Hz
*RST: 0 Hz
Default unit: Hz
Example:
FREQ:STAR 2GHz; STOP 3 GHz
Define a sweep range between 2 GHz and 3 GHz.
FREQ:OFFS:PWAV 10kHz; WAV 20kHz
Modify the receiver frequency settings.
[SENSe<Ch>:]FREQuency<Pt>:OFFSet:WAVes <FreqOffset>
Defines the constant receiver frequency offset for unprimed wave quantities a, b rela-
tive to the common receiver frequency.
Suffix:
<Ch>
.
Channel number.
<Pt> Test port number of the analyzer.
Parameters:
<FreqOffset> Frequency offset.
Range: 0 Hz to 25 MHz
Increment: 1 Hz
*RST: 0 Hz
Default unit: Hz