Command reference
R&S
®
ZNA
1567User Manual 1178.6462.02 ─ 20
Tip: This command is complemented by CALCulate<Chn>:LIMit:RDOMain:
COMPlex and CALCulate<Chn>:LIMit:RDOMain:FORMat.
Suffix:
<Chn>
.
Channel number used to identify the active trace.
Setting parameters:
<Format> LINear | LOGarithmic | DB | SIC
Keyword for the physical unit of the response values.
Default unit: dB (irrespective of the parameter selected)
Usage: Setting only
CALCulate<Chn>:LIMit:LOWer:STATe <Boolean>
CALCulate<Chn>:LIMit:UPPer:STATe <Boolean>
These commands switch the lower and upper limit check on or off. Lower limit line seg-
ments are assigned even numbers; upper limit line segments are assigned odd num-
bers; see CALCulate<Chn>:LIMit:LOWer[:DATA] and CALCulate<Chn>:
LIMit:UPPer[:DATA]. CALCulate<Chn>:LIMit:LOWer:STATe does not affect
segments with odd numbers; CALCulate<Chn>:LIMit:UPPer:STATe does not
affect segments with even numbers.
Note: Use CALCulate<Chn>:LIMit:STATe to switch on or off the entire limit check,
including upper and lower limit lines.
Suffix:
<Chn>
.
Channel number used to identify the active trace.
Parameters:
<Boolean> ON | OFF - switch limit check on or off.
*RST: OFF
Example:
CALC:LIM:LOW -10, 0, 0, -10
Define two limit line segments covering the entire sweep range.
Two upper limit line segments with default response values are
created in addition.
CALC:LIM:UPP 0, 5, 5, 0
Change the response values of the upper limit line segments.
CALC:LIM:LOW:STAT ON; :CALC:LIM:UPP:STAT ON; :
CALC:LIM:FAIL?
Switch the limit check on and query the result.
Manual operation: See "Limit Check" on page 432
CALCulate<Ch>:PARameter:DEFine <TraceName>, <Result>[, <TestPortNum>]
Creates a trace and assigns a channel number, a name and a measurement parame-
ter to it. The trace is not displayed. To display a trace defined via
CALCulate<Ch>:PARameter:DEFine, a window must be created (DISPlay[:
WINDow<Wnd>][:STATe] ON) and the trace must be assigned to this window
(DISPlay[:WINDow<Wnd>]:TRACe<WndTr>:FEED); see example below.
R&S ZVR/ZVABT compatible commands