Traces must be selected to become active traces; see CALCulate<Ch>:PARameter:
SELect.
Note: The parameter names in this command differ from R&S ZNA conventions; more-
over the parameter list is not complete. The alternative command CALCulate<Ch>:
PARameter:SDEFine uses a complete parameter list with compatible names.
Suffix:
<Ch>
.
Channel number. <Ch> may be used to reference a previously
defined channel. If <Ch> does not exist, it is generated with
default channel settings.
Setting parameters:
<TraceName> Trace name, e.g. 'Trc4'. See "Rules for trace names" in "Table
Area" on page 390.
<Result> S11 | S12 | S13 | S14 | S21 | S22 | S23 | S24 | S31 | S32 | S33 |
S34 | S41 | S42 | S43 | S44 | A | B | C | D | R1 | R2 | R3 | R4 |
AB | AC | AD | BA | BC | BD | CA | CB | CD | DA | DB | DC |
AR1 | AR2 | AR3 | AR4 | BR1 | BR2 | BR3 | BR4 | CR1 | CR2 |
CR3 | CR4 | DR1 | DR2 | DR3 | DR4 | R1A | R1B | R1C | R1D |
R2A | R2B | R2C | R2D | R3A | R3B | R3C | R3D | R4A | R4B |
R4C | R4D | R1R2 | R1R3 | R1R4 | R2R1 | R2R3 | R2R4 |
R3R1 | R3R2 | R3R4 | R4R1 | R4R2 | R4R3
Measurement parameter; see list of parameters below.
<TestPortNum> Test port number, drive port for wave quantities and ratios,
ignored for S-parameters.
Example:
CALC4:PAR:DEF 'Ch4Tr1', S11
Create channel 4 and a trace named Ch4Tr1 to measure the
input reflection coefficient S11.
DISP:WIND:STAT ON
Create diagram area no. 1.
DISP:WIND:TRAC2:FEED 'CH4TR1'
Display the generated trace in diagram area no. 1, assigning a
trace number 2.
Usage: Setting only
The measurement parameter is selected by means of the following keywords (the
selection depends on the number of test ports of the analyzer, e.g. S44 is not available
on 2-port analyzers):
S11 | S12 | S13 | S14 | S21 | S22 | S23 | S24 | S31 |
S32 | S33 | S34 | S41 | S42 | S43 | S44
S-parameters
A | B | C | D Wave quantities b
1
, b
2
, b
3
, b
4
(received waves)
R1 | R2 | R3 | R4 Wave quantities a
1
, a
2
, a
3
, a
4
(reference waves)
AB | AC | AD | BA | BC | BD | CA | CB | CD | DA | dB
| DC
Ratio of wave quantities b
1
/b
2
, b
1
/b
3
, ... , b
4
/b
3
(received waves only)