GUI reference
R&S
®
ZNA
433User Manual 1178.6462.02 ─ 20
Note:
Limit check and display of limit lines are independent of each other:
The limit lines can be displayed, no matter if the limit check is enabled.
If "Limit Check" is enabled, the limits are checked, no matter if the limit lines
are displayed.
The limit check can even be enabled, if no limit lines are defined. In this case,
the info field displays "No limit defined!" and the limit check always passes.
Limit lines are defined for a particular trace format. However, the limit check is per-
formed irrespective of the current trace format. The info field indicates the correct
"PASS"/"FAIL" state and limit violations are visualized on the trace (if any).
Figure 5-15: Limit line violations in complex trace formats
Remote command:
CALCulate<Chn>:LIMit:STATe
CALCulate<Chn>:LIMit:LOWer:STATe
CALCulate<Chn>:LIMit:UPPer:STATe
CALCulate<Chn>:LIMit:FAIL?
CALCulate:LIMit:FAIL:ALL?
CALCulate<Chn>:LIMit:STATe:AREA
Limit Fail Beep
Activates or deactivates the fail beep. The fail beep is a low-tone acoustic signal that is
generated each time the analyzer detects an exceeded limit. No fail beep can be gen-
erated if the limit check is switched off.
Note: The R&S ZNA does not have a built-in audio device and loudspeaker. To hear
these sounds, connect a USB audio device to the instrument or operate it via remote
desktop.
Remote command:
CALCulate<Chn>:LIMit:SOUNd[:STATe]
Clear Test
Resets the limit check results.
Remote command:
CALCulate<Chn>:LIMit:CLEar
Lines softtool