GUI reference
R&S
®
ZNA
516User Manual 1178.6462.02 ─ 20
Figure 5-20: Select Calibration Properties dialog: scalar mixer measurements
For the standard mode a calibration unit is used, if connected.
Measurement
mode
Traces Power meter Calibration setup
Standard only one-port OSM
POSM
at least one two-
port
UOSM
PUOSM
Spectrum
(R&S ZNA-K1)
(required) POSM
Scalar mixer
(R&S ZNA-K4)
(required) 1x PUOSM
or
2x POSM
*
(see screenshot above)
IMD (R&S ZNA-K4)
(required) "P-Ref Receiver" for UT and LT port, if on
VNA,
and
"P-Trans Norm" between UT and IF port
Vector mixer
(R&S ZNA-K5)
"PUOSM Scalar Vector Mixer"
or
"UOSM Vector Mixer"
*
automatically selected if frequency converters are involved
Cal softtool