EasyManua.ls Logo

R&S ZNA26 - Page 557

R&S ZNA26
1709 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
GUI reference
R&S
®
ZNA
557User Manual 1178.6462.02 ─ 20
Use "Detect Port Assignment" to detect the current port assignment (after reconnecting
the calibration unit).
Measurement
Selects the measurement to be performed in advanced mode and runs it
Characterization
Selects the calibration unit characterization to be used for the validation.
By default the factory characterization is used.
Remote command:
CALCulate:CALValidate:CHARacteriza
Calibration Standard
Selects the calibration standard to be measured in advanced mode.
Validation Limits
The "Limits..." button opens the "Cal Validation Limits" dialog that allows you to specify
the maximum tolerable deviations from the corresponding S-parameters of the
selected reference characterization.
Defaults are ±0.2 dB for magnitude and ±2° for phase deviations.
Remote command:
CALCulate:CALValidate:RESPonse:MAGNitude:LOWer
CALCulate:CALValidate:RESPonse:MAGNitude:UPPer
CALCulate:CALValidate:RESPonse:PHASe:LOWer
CALCulate:CALValidate:RESPonse:PHASe:UPPer
Cal softtool

Table of Contents

Other manuals for R&S ZNA26

Related product manuals