GUI reference
R&S
®
ZNA
632User Manual 1178.6462.02 ─ 20
5.13.1.1 Controls on the Mode tab
Driving Mode
Determines the order of partial measurements and sweeps.
In "Auto" mode, the analyzer optimizes the display update: Fast sweeps are per-
formed in "Alternated" mode, slower sweeps in "Chopped" mode.
In "Alternated" mode, a partial measurement is performed at all sweep points (par-
tial sweep) before the hardware settings are changed. The next partial measure-
ment is carried out in an additional sweep.
This mode is usually faster than "Chopped" mode.
In "Chopped" mode, the analyzer completes the necessary sequence of partial
measurements at each sweep point and obtains the result (measurement point)
before proceeding to the next sweep point. A trace is obtained from the beginning
of the sweep.
The "Driving Mode" setting is also used during a system error correction. For channels
which require a single partial measurement only, the driving mode settings are equiva-
lent. See also Chapter 4.1.4.1, "Partial measurements and driving mode", on page 88.
Remote command:
[SENSe<Ch>:]COUPle
Image Suppr.
The "Image Suppr." settings define whether the analyzer measures with a local oscilla-
tor frequency LO below or above the RF input frequency. This feature can be used to
Mode softtool