GUI reference
R&S
®
ZNA
833User Manual 1178.6462.02 ─ 20
Cable & Splitter Loss
Defines the loss of
the cables for source ports
the cables and the splitter (if used) for LO ports
With Apply to all channels in all setups (or after a [Preset]), these values are written to
the Port Power Offset of the respective ports.
Remote command:
[SENSe:]CONVerter<Port>:RFLoss
[SENSe:]CONVerter<Port>:CSLoss
Slope
Defines the slope factor of cabling (and splitter).
With Apply to all channels in all setups (or after a [Preset]), these values are written to
the Slope of the respective ports.
Remote command:
[SENSe:]CONVerter<Port>:CSSLope
Apply to all channels in all setups
Use "Apply to all channels in all setups" to write the Cable & Splitter Loss and Slope
values to the Port Power Offset and Slope channel settings of the respective ports.
Remote command:
SOURce<Ch>:POWer<PhyPt>[:LEVel][:IMMediate]:OFFSet
SOURce<Ch>:POWer<PhyPt>:GENerator<Gen>:OFFSet
SOURce<Ch>:POWer<PhyPt>[:LEVel][:IMMediate]:SLOPe
5.19.6.4 Leveling Datasets dialog
After you went through the Leveling procedure, you can configure the usage of the
available leveling datasets, persist and recall them.