GUI reference
R&S
®
ZNA
837User Manual 1178.6462.02 ─ 20
Access: System – [Setup] > "Generic Device" > "Generic Device..."
Background information
Refer to Chapter 4.7.40, "Generic devices", on page 292.
The configuration of a new generic device involves the following steps:
1. Connect the device to your R&S ZNA using a LAN (VXI-11), GPIB, or USB inter-
face.
2. If the device is connected via LAN, enable "LAN detection"
3. Select "Scan Instruments" and wait until the device appears in the table of "Known
Devices".
4. Select "+" to copy the device to the list of configured devices.
If the R&S ZNA fails to detect a connected external device, click Add Device to define
the interface type and address manually.
To control external devices via GPIB, the pre-installed RS Visa library (visa32.dll)
must be replaced by the NI Visa library.
Known Devices/Address Filter/Clear
Table with all VISA-capable devices that the analyzer detects to be online (i.e. connec-
ted and switched on). Scan Instruments refreshes the table. The R&S ZNA can auto-
detect the serial number of the connected devices.
"Address Filter" allows you to filter the known devices to the specified address frag-
ment, "Clear" removes the filter.
Setup softtool