Command reference
R&S
®
ZNA
935User Manual 1178.6462.02 ─ 20
Suffix:
<Ph_pt>
.
Physical port number
Parameters:
<Boolean> ON (1)
Port <Ph_pt> is active in the respective measurement
OFF (0)
Port <Ph_pt> is inactive in the respective measurement
Options: R&S ZNA-K231
Manual operation: See "Ports table" on page 703
CALCulate:FMODel:DELT:M1L:DIFFmode
<Boolean>
The one-length method (CALCulate:FMODel:DELT<Ph_pt>:METHod M1L) handles
single-ended and balanced measurement results differently. This command selects
between the two modes.
Parameters:
<Boolean> OFF (0)
Single-ended mode
ON (1)
Differential mode
*RST: OFF (0)
Options: R&S ZNA-K231
Manual operation: See "Method" on page 700
CALCulate:FMODel:DELT:MEASurement <Measurement>
Defines whether single-ended or balanced ports are measured for Delta-L characteri-
zations.
Parameters:
<Measurement> SINGleended | BALanced
SINGleended
Single-ended ports only
BALanced
Balanced ports only
*RST: SINGleended
Options: R&S ZNA-K231
Manual operation: See "Measurements" on page 700
CALCulate:FMODel:DELT<Ph_pt>:METHod <Method>
Selects the Delta-L+ characterization method to be used.
SCPI command reference