Command reference
R&S
®
ZNA
937User Manual 1178.6462.02 ─ 20
Suffix:
<Ph_pt>
.
This suffix is ignored
Parameters:
<SweepControl> DEFault | CURRent
DEFault
Use the default values of the Delta-L algorithm:
10 MHz frequency step size, 1 kHz IF bandwidth
SweepControl
Use the values of the active channel
Options: R&S ZNA-K231
Manual operation: See "Sweep Control" on page 701
CALCulate:FMODel:DELT<Ph_pt>:SWEep:CONTrol:BWIDth[:RESolution]:
DREDuction <Boolean>
Defines whether dynamic bandwidth at low frequencies is used for Delta-L measure-
ments (see [SENSe<Ch>:]BANDwidth[:RESolution]:DREDuction)
Suffix:
<Ph_pt>
.
This suffix is ignored
Parameters:
<Boolean> ON (1)
Dynamic bandwidth at low frequencies is active (default)
OFF (0)
Dynamic bandwidth at low frequencies is inactive
Options: R&S ZNA-K231
Manual operation: See "Sweep Control" on page 701
CALCulate:FMODel:DELT<Ph_pt>:SWEep:CONTrol:IFBW?
For the Delta-L measurement, the firmware can either use a default IF bandwidth, or
the active channel setting (see CALCulate:FMODel:DELT<Ph_pt>:SWEep:
CONTrol).
This query returns the resulting IF bandwidth.
Suffix:
<Ph_pt>
.
This suffix is ignored
Usage: Query only
Options: R&S ZNA-K231
Manual operation: See "Sweep Control" on page 701
SCPI command reference