Command reference
R&S
®
ZNA
940User Manual 1178.6462.02 ─ 20
Parameters:
<String> Path to the Touchstone file, either absolute or relative to the
selected working directory (CALCulate:FMODel:DIRectory).
Options: R&S ZNA-K210
Manual operation: See "Load File / 1x Open Preset / 1x Short Preset" on page 677
CALCulate:FMODel:EZD<Ph_pt>:COUPon:MEASure:FILename:CLEar
Invalidates the file path previously set using CALCulate:FMODel:EZD<Ph_pt>:
COUPon:MEASure:FILename.
Suffix:
<Ph_pt>
.
This suffix is ignored.
Usage: Event
Options: R&S ZNA-K210
CALCulate:FMODel:EZD<Ph_pt>:COUPon[:STATe]
<Boolean>
For a fixture modeling with the EZD tool, this command allows to specify the ports to
which the test coupon is connected.
This has to be defined before measuring the test coupon (using CALCulate:FMODel:
EZD<Ph_pt>:COUPon:MEASure).
Suffix:
<Ph_pt>
.
Physical port number
Parameters:
<Boolean> ON (1)
if the test coupon is connected to port <Ph_pt>
OFF (0)
otherwise
Options: R&S ZNA-K210
Manual operation: See "Active" on page 677
CALCulate:FMODel:EZD<Ph_pt>:DCEXtrapolat <Boolean>
Defines whether the Touchstone de-embedding file generated by the EZD tool contains
data for DC.
Suffix:
<Ph_pt>
.
This suffix is ignored.
Parameters:
<Boolean> ON (1)
The EZD tool adds (extrapolated) S-parameter data for f = 0.
SCPI command reference