Preparing the Analyzer for Use
R&S
®
ZVA
29Getting Started 1145.1090.62 ─ 13
For detailed information about Remote Desktop refer to the Windows
®
Help.
1.9 Firmware Update
Upgrade versions of the analyzer firmware are supplied as single setup files *.msi. To
perform a firmware update,
1. Copy the setup file to any storage medium accessible from the analyzer. This may
be the internal hard disk, an external storage medium (USB memory stick, CD-
ROM with external drive) or a network connection (LAN, GPIB bus).
2. Double-click the setup file (or use the front panel keys to select and start the setup
file; see chapter 2.3.2.1, "Using Front Panel Keys", on page 39) and follow the
instructions of the setup wizard.
Setup files can be stored and installed again. The default drive letter of the internal
hard disk is "C". External storage devices are automatically mapped to the next free
drive letters "D", "E" etc.
Factory calibration
A firmware update does not affect the factory calibration.
External accessories
Calibration units and extension Units must be disconnected during a firmware update.
Firmware Update