System Overview
R&S
®
ZVA
89Getting Started 1145.1090.62 ─ 13
Imported and user-defined kits can be changed in the "Calibration Kits" dialog and
its various sub-dialogs.
Calibration kits and connector types are global resources; the parameters are stored
independently and are available irrespective of the current setup.
3.4.2 Calibration Types
The analyzer provides a wide range of calibration types for one, two or more ports. The
calibration types differ in the number and types of standards used, the error terms, i.e.
the type of systematic errors corrected and the general accuracy. The following table
gives an overview.
Table 3-2: Overview of calibration types
Calibration Type Standards Parameters Error Terms General Accuracy Application
Reflection Normali-
zation
Open or Short S
11
(or S
22
, ...)
Reflection tracking Low to medium Reflection measure-
ments on any port.
Transmission Nor-
malization
Through S
12
, S
21
(or S
13
,...)
Transmission track-
ing
Medium Transmission mea-
surements in any
direction and
between any combi-
nation of ports.
Full One-Port Open, Short and
Match
1)
S
11
(or S
22
, ...)
Reflection tracking,
Source match
Directivity,
High Reflection measure-
ments on any port.
One-Path Two-Port Open, Short and
Match
1)
(at source
port),
Through
2)
S
11
, S
21
(or S
22
,...)
Reflection tracking,
Source match,
Directivity,
Transmission track-
ing
Medium to high Unidirectional trans-
mission measure-
ments in any direc-
tion and between
any combination of
ports.
TOSM (2 or more
ports) or UOSM
Open, Short,
Match
1)
at all ports,
Through
2)
between
all directed port
pairs
All Reflection tracking,
Source match,
Directivity,
Load match,
Transmission track-
ing,
High Reflection and
transmission mea-
surements on DUTs
with 2 or more
ports; classical 12-
term error correc-
tion model.
TOM (2 or more
ports)
Open and Match at
all ports,
Through between
all directed port
pairs
All Reflection tracking,
Source match,
Directivity,
Load match,
Transmission track-
ing
High, implicit verifi-
cation
Reflection and
transmission mea-
surements on DUTs
with 2 or more
ports.
Calibration

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the R&S ZVA Series and is the answer not in the manual?

R&S ZVA Series Specifications

General IconGeneral
BrandR&S
ModelZVA Series
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals