47
Appendix B
46
Appendix B
STANDARD MODEL
Optical Chart
3000
120
1.0
0
0
24
SPOT DIA. (IN)
SPOT DIA. (mm)
DISTANCE: SENSOR TO OBJECT (mm)
DISTANCE: SENSOR TO OBJECT (IN)
FOCUS POINT D:S = 60:1 FAR FIELD D:S = 35:1
0.76 IN @ 46 IN
19 mm @ 1150 mm
1500
2500
100
60
1.2
2.3
2.9
29.0
58.0
72.0
2000
44.0
1000
500
250
100
24
23
30
0.9
0.81
0.82
1.51
1.85
2.5
19.8
108
84
72
48
20
36
46
0.87
The measured spot size depends on the
distance between the object you are measuring
and the infrared thermometer.
The relationship between distance and spot size
is 60:1 at the focus point.
The D:S in the far field (>33ft/10m) is 35:1.
CLOSE FOCUS MODEL
Optical Chart
500
20
1
0
0
0.93
25
23
SPOT DIA. (IN)
SPOT DIA. (mm)
DISTANCE: SENSOR TO OBJECT (mm)
DISTANCE: SENSOR TO OBJECT (IN)
FOCUS POINT D:S = 50:1 FAR FIELD D:S = 12:1
0.24 IN @ 11.8 IN
6 mm @ 300 mm
Close Focus
6
0.62
15,5
150
2
0.9
22
50
4
0.75
18,7
100
7.9
9.8
200
250
0.49
0.37
12,3
9,2
40
1000
2.81
70
The measured spot size depends on the distance
between the object you are measuring and the
infrared thermometer. The relationship between dis-
tance and spot size is 50:1 at the focus point.
The D:S in the far field is 12:1.
The best distance between sensor and object is from
10 cm (4“) up to 50 cm (20“).
Optical Resolution (Infrared Themometer)
Standard Focus Model
Optical Resolution (Infrared Thermometer)