9 to 36 V DC 250 mA max (Isolated from output)
4-20mA into 100/550 max. Overload internally limited to
30mA max. This is an active output that should not be
connected to any external power supply as this will
damage unit.
1V rms. at 5kHz standard. 25mA maximum.
3Vrms and 3 kHz also available.
Oscillator Temperature Coefficient
X0.07 to x500 in 8 ranges with fine control interpolation
Zero Stability Voltage Output
Current Output
0.002% of FS typical/°C (optimum at ±10V o/p)
0.005% of FS typical/°C
Gain Stability Voltage Output
Current Output
0.005% of FS typical/°C (optimum at ±10V o/p)
0.01% of FS typical /°C
Noise -Voltage Output
Current Output
Current Output
When subjected to radiated electro-magnetic energy (as
EN61000-4-3) an additional error can occur at certain
frequencies: