EasyManua.ls Logo

RDS CMM 100 - Calibration Basics

RDS CMM 100
23 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
CMM 100 : INSTALLATION, CALIBRATION AND OPERATION
5
The operator is able to select from channels "A" to "F", which correspond to
different crop types.
The head unit is normally powered on via the ignition circuit. It has a rear-
mounted on-off switch that also enables entry into the calibration mode
without having to switch the engine off.
1.2 Calibration
The grain moisture value should still be checked periodically against a
reading from a reference moisture meter, particularly when combining oily or
dirty crops, since the sensor can produce a false reading due to extraneous
material adhering to it.
The CMM 100 has factory-default calibration settings for each channel A to F.
These settings assume channels A to F are assigned to the following crop
types,
A = Wheat
B = Barley
C= Oats
D = Oilseed Rape
E = Beans
F = Maize
Adjustment is needed if the harvest moisture reading is different from that of
a reference moisture meter.
The operator can adjust the calibration settings for each channel, including:
Calibration Factors "A", "B" and "C" for calibrating the sensor.
Excess Moisture Content - Alarm Threshold.
Low Grain Voltage Threshold for when there is insufficient grain passing
over the sensor to give a sensible reading of moisture content, such as
occurs when the combine harvester /drier first starts up. By default, this
feature is switched off in the CAL mode ("UOLt" setting).
The operator changes the calibration settings for each channel using a
secure method of entry that helps to prevent any accidental changes.
Please refer to the "Calibration" section.

Related product manuals