CMM 100 : INSTALLATION, CALIBRATION AND OPERATION
9
2.6.2 Set High/ Low Moisture Content Alarm Threshold
Having first set the "HIGH / "LO" alarm option (section 2.6.1), you can then
individually set the alarm threshold (% M.C.) for each channel as desired.
1. In CAL mode press the button to cycle through the factors until "ALAr"
is displayed. The default threshold is displayed for the selected channel.
2. Press the button to select the desired channel.
3. Edit the setting as described in section 3.2.
NOTE: Changing the "HIGH" / "LO" alarm option resets any pre-programmed
threshold values to the factory default values (25% / 5%).