EasyManua.ls Logo

red lion IAMS0002 - Input Specifications; Basic Input Values; RTD and Potentiometer Inputs

red lion IAMS0002
40 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
LP1109 15
Basic values
Input
type
Basic
accuracy
Temperature
coefficient
mA ±4 µA ±0.4 µA / °C
Volt ±20 µV ±2 µV / °C
Pt100 ±0.2°C ±0.01°C / °C
Linear resistance ±0.1 Ω ±0.01 Ω / °C
Potentiometer ±0.1 Ω ±0.01 Ω / °C
TC type:
E, J, K, L, N, T, U
±1°C
±0.05°C / °C
TC type: R, S, W3,
W5, LR
±2°C
±0.2°C / °C
TC type: B
85...200°C
±4°C
±0.4°C / °C
TC type: B
200...1820°C
±2°C
±0.2°C / °C
Auxiliary supplies:
2-wire supply (terminal 44...43) ........................ 25...16 VDC / 0...20 mA
RTD, linear resistance and potentiometer input
Input for RTD types:
Pt10, Pt20, Pt50, Pt100, Pt200, Pt250, Pt300, Pt400, Pt500, Pt1000
Ni50, Ni100, Ni120, Ni1000, Cu10, Cu20, Cu50, Cu100
Input
type
Min.
value
Max.
value
Standard
Pt10...Pt1000
Ni50...Ni1000
Cu10...Cu100
Lin. R
Potentiometer
-200°C
-60°C
-200°C
0 Ω
10 Ω
+850°C
+250°C
+260°C
10000 Ω
100 kΩ
IEC 60751
DIN 43760
α = 0,00427
-
-
Cable resistance per wire (max.), RTD............. 50 Ω
Sensor current, RTD ................................................ Nom. 0.2 mA
Effect of sensor cable resistance
(3- / 4-wire), RTD ..................................................... < 0.002 Ω / Ω
Sensor error detection, RTD ................................ Ye s
Short circuit detection, RTD ................................ < 15 Ω
EMC immunity influence .................................................. < ±0.5% of span
Extended EMC immunity:
NAMUR NE 21, A criterion, burst ................................. < ±1% of span