16 LP1109
TC input
Type
Min.
value
Max.
value
Standard
B
E
J
K
L
N
R
S
T
U
W3
W5
LR
0°C
-100°C
-100°C
-180°C
-200°C
-180°C
-50°C
-50°C
-200°C
-200°C
0°C
0°C
-200°C
+1820°C
+1000°C
+1200°C
+1372°C
+900°C
+1300°C
+1760°C
+1760°C
+400°C
+600°C
+2300°C
+2300°C
+800°C
IEC 60584-1
IEC 60584-1
IEC 60584-1
IEC 60584-1
DIN 43710
IEC 60584-1
IEC 60584-1
IEC 60584-1
IEC 60584-1
DIN 43710
ASTM E988-90
ASTM E988-90
GOST 3044-84
Cold junction compensation (CJC)
via internal CJC sensor ...................................... ±(2.0°C + 0.4°C * Δt)
Δt = internal temperature - ambient temperature
Sensor error detection, all TC types................. Yes
Sensor error current:
when detecting .................................................... Nom. 2 μA
else ........................................................................... 0 μA
Current input
Measurement range ................................................ 0...20 mA
Programmable measurement ranges ............... 0...20 and 4...20 mA
Input resistance........................................................ Nom. 20 Ω + PTC 50 Ω
Sensor error detection:
Loop break 4...20 mA ......................................... Yes
Voltage input
Measurement range ................................................ 0...12 VDC
Programmable measurement ranges ............... 0...1 / 0.2...1 / 0...5 / 1...5 /
0...10 and 2...10 VDC
Input resistance........................................................ Nom. 10 MΩ