ERROR INDICATORS
Air plug disconnected.
Arm cuff not applied
correctly.
Air is leaking from the
arm cuff.
Cuff over or under
inflated.
Device error.
Batteries are worn.
Movement during
measurement.
Clothing is interfering
with the arm cuff.
Insert the plug securely. Refer to
“Applying the Arm Cuff”.
Apply the arm cuff correctly. Refer to
“Applying the Arm Cuff”.
Do not inflate the cuff above 299 mmHg.
Refer to “Taking a Measurement”.
Remove any clothing interfering with the
arm cuff. Refer to “Applying the Arm
Cuff”.
Replace the four batteries.
Refer to “Battery Installation”.
Repeat measurement. Remain still and do
not talk during measurement. Refer to
“Taking a Measurement”.
Repeat measurement. Remain still and do
not talk during measurement. Refer to
“Taking a Measurement”.
If “E2” appears repeatedly, inflate the cuff
manually until it is 30 to 40 mmHg above
your previous measurement result. Refer
to “Taking a Measurement”.
Replace the cuff with a new one.
Contact Customer Service.
The arm cuff was inflated
above 299 mmHg when
inflating the cuff manually.
SYMBOL CAUSE CORRECTION