Details FCC and RoHS compliance, and packaging recyclability.
Outlines conditions for FCC compliance and potential interference.
Confirms compliance with EC directive 2011/65/EU.
States disclaimer and directs users to additional documentation.
Lists compatible and incompatible solvents for cleaning and handling.
Specifies temperature and humidity ranges for storage and operation.
Details minimum bend radius for the scale during installation and handling.
Provides detailed dimensional data for readhead mounting and alignment.
Illustrates positions for limit switch and reference mark actuators.
Shows required clearance between readhead and scale surface.
Shows overall and scale lengths, and actuator position ranges.
Notes requirements for scale mounting surface roughness and parallelism.
Details dimensions for an optional bolted reference mark actuator.
Explains how to use scale applicators for installation.
Details the procedure for applying the scale to the substrate.
Describes fitting end clamps for scale stability and repeatability.
Guides on positioning actuators using the orange shim.
Lists different size limit switch actuators.
Shows options for epoxy-mounted and screw-mounted actuators.
Specifies criteria for readhead mounting brackets for stability and adjustment.
Details using spacers and checking the LED for correct rideheight and alignment.
Explains phasing the reference mark for consistent datuming.
Describes the process using the set-up LED and adjuster screw.
Details quadrature signals for different models.
Explains Z signal duration and limit switch outputs.
Describes alarm indication and external set-up signal usage.
Details analogue V, V2, and V0 signals.
Explains BID/DIR pin functions for reference mark direction.
Lists maximum speeds and counter input frequencies for readhead types.
Emphasizes proper connection of shields to machine earth and 0V.
Shows recommended termination for digital outputs with capacitors.
Details termination for analogue V, V2, V0 and limit outputs.
Defines signal period, resolution, and phase for digital outputs.
Explains Z signal duration and limit switch outputs.
Details conditions that trigger alarm outputs for different readhead types.
Describes sinusoidal analogue incremental signals.
Details analogue reference pulse and limit output signals.
Explains the Vx signal for set-up purposes.
Lists power supply, ripple, and temperature limits.
Details humidity, sealing (IP50), and resistance to acceleration and vibration.
Provides mass of readhead and cable details including flex life.
Describes RGS20-S and RGS20-P scale types and their period.
Specifies linearity tolerances and available scale lengths.
Details scale dimensions, substrate compatibility, and end fixing methods.