J
Jamie SantosSep 14, 2025
How to fix not getting a green LED over the complete axis length on Renishaw Media Converter?
- Ccarol60Sep 15, 2025
If you're unable to get a green LED over the complete axis length, it indicates that the system run-out is not within specification. To fix this: * Check that the readhead variant is the correct type for the chosen scale. * Use a DTi gauge to check that the run-out is within specifications. * Restore factory defaults. * Realign the readhead to obtain a green LED at the mid-point of the run-out.