English 1
Disinfection guide Clinical use only
This guide is intended for multi-patient re-use of the AirFit N30i and AirFit P30i masks in a sleep lab, clinic or hospital. If you use the mask as a single user in the home, refer to the User Guide for
cleaning instructions. The processor is responsible for ensuring that reprocessing is completed in accordance with ResMed’s validated procedures.
Mask component
1
Cleaning - Alconox
Validated number of cycles
Thermal Disinfection - Manual
167°F (75°C) for 30 minutes
194-199°F (90-93°C) for 1 to 10 minutes
• AirFit N30i cushion
30
• AirFit P30i cushion
30
• Multi-hole vent
30
• Frame (SLM)
2
30
• Elbow body / elbow clip / elbow ring
30
• Headgear
30
1
This mask may not be available in all regions. For full details regarding the correct use of this mask, please refer to the specific User Guide. For a list of available replacement parts for each mask system, check the Product Guide on ResMed.com.
2
The frame of the Sleep Lab Mask (SLM) does not contain sleeves.
A Elbow body
B Elbow ring
C Headgear
D Frame (SLM)
E AirFit N30i nasal cushion
F AirFit P30i pillows cushion
G Elbow clip
1 Vent
2 Multi-hole vent (AirFit P30i only)