TROUBLESHOOTING
ENGLISH
TROUBLESHOOTING
!
CAUTION
If the problem cannot be solved, contact your equipment supplier or ResMed. Do
not attempt to open the S8 Compact case. There are no user serviceable parts
inside. Repairs and internal servicing should only be performed by an authorized
service agent.
Problem Possible Cause Solution
No display. Power is not connected. Ensure the power cable is
connected and the power
switch (if available) is on.
Treatment pressure
seems low.
Ramp Time is in use. Wait for air pressure to build
up or change ramp time.
Air filter is dirty. Replace air filter.
Air tubing is kinked or
punctured.
Straighten or replace tubing.
Air tubing is not connected
properly.
Check air tubing.
Mask and headgear are not
positioned correctly.
Adjust position of mask and
headgear.
Caps are missing from access
port on mask.
Replace caps.
Altitude compensation setting is
incorrect.
Adjust altitude
compensation setting to the
correct altitude range.
Pressure required for treatment
may have changed.
Calibrate pressure.
Excessive air delivered
from the S8 Compact.
Altitude compensation setting is
incorrect.
Adjust altitude
compensation setting to the
correct altitude range.