Operating temperature +5ºC to +35ºC
Operating humidity 10–95% non-condensing
Operating altitude Sea level to 2,591 m
Storage and transport temperature -20ºC to +60ºC
Storage and transport humidity 10–95% non-condensing
Housing construction Flame retardant engineering thermoplastic
Supplemental oxygen Recommended maximum supplemental oxygen flow:
4 L/min
Hypoallergenic air filter Non-woven acrylic and polypropylene fibers with
polypropylene carrier
Standard air filter Polyester fiber
SlimLine
™
air tubing Flexible plastic, 1.8 m, 15 mm inner diameter
Standard air tubing Flexible plastic, 2 m, 19 mm inner diameter
ClimateLine
™
heated air tubing Flexible plastic and electrical components, 2 m, 15 mm
inner diameter
ClimateLine
MAX™
heated air tubing Flexible plastic and electrical components,
1.9 m, 19 mm inner diameter
Air outlet The 22 mm conical air outlet complies with ISO 5356-1
Electromagnetic compatibility Product complies with all applicable electromagnetic
compatibility requirements (EMC) according to
IEC60601-1-2, for residential, commercial, and light industry
environments. Information regarding the electromagnetic
emissions and immunity of these ResMed devices can be
found on www.resmed.com, on the Products page under
Service and Support. Click on the PDF file for your language.
Aircraft use Medical-Portable Electronic Devices (M-PED) that meet the
Federal Aviation Administration (FAA) requirements of
RTCA/DO-160F can be used during all phases of air travel
without further testing or approval by the airline operator.
ResMed confirms that the S9 meets RTCA/DO-160F
requirements.
IEC 60601-1 classification Class II (double insulation), Type BF
368147 S9 AutoSet S9 Elite User Guide ROW.book Page 7 Thursday, August 4, 2011 8:57 AM