Operating Temperature | 5°C to 35°C (41°F to 95°F) |
---|---|
IP Rating | IP21 |
Device Type | Ventilator |
Modes of Operation | CPAP, S, T, ST, PAC, iVAPS |
Weight | 2.1 kg (device with power supply) |
Power Supply | 100-240V AC, 50/60Hz |
Battery Backup | Up to 2 hours with internal battery |
Operating Humidity | 10% to 95% relative humidity, non-condensing |
Sound Level | 30 dBA |
Data Storage | SD card |
Alarms | High pressure, Low pressure, Power failure, Disconnection, Low battery |
Target Tidal Volume | Adjustable |
Keep battery connected to charger and plugged into grounded outlet. Disconnect battery before device connection.
Locate the white Hubbell connector on the battery and the black connector on the charger.
Twist-lock connectors together. Plug charger into a properly grounded 3-prong electrical outlet.
Yellow LED flashes during charging. Green LED indicates charged. Charger enters hibernation mode.
Green AC LED indicates power to the Stellar™150 under normal operation.
Battery LED illuminates when using external battery. 'Internal Battery in Use!' message shown.
Alarm LED and tone indicate power loss. Battery icon shows charge level.
Assemble cables if needed. Connect black Hubbell connector to white Hubbell connector clockwise.
Ensure auxiliary power connector toggle switch is in the ON (I) position.
Insert connector firmly into the DC power inlet at the back of the device.
Alerts for 30% and 15% internal battery capacity. Connect external battery if needed.
Requires mains power or external battery. Restart therapy using the main power switch.
Disconnect external battery when mains power returns. Reconnect to charger.