24
ReSound BTEs
25
To dispose of the used wax fi lter:
1. With the used wax fi lter attached to the HF3
tool, insert the tool into the center slot of the fi lter
package.
2. Slide the tool to the side. The used wax fi lter
will remain in the slot when you lift the HF3 tool
straight out.
To install a new wax fi lter:
1. Press down on a new wax fi lter in the package
with the plastic end of the HF3 tool.
2. Lift straight up. A new wax fi lter will cling to the
end of the HF3 tool.
3. Press the new wax fi lter straight into the tip of the
receiver.
4. Gently move the HF3 tool side to side while pull-
ing it away from the hearing instrument. The new
wax fi lter is now installed.
HF3 WAX PROTECTION SYSTEM
(RIE MODELS ONLY)
An HF3 wax fi lter is mounted in front of the receiver
tube for protection against wax and moisture. Use
the small cleaning brush to gently brush away loose
wax and debris from the tip of the
receiver of the hearing instrument.
This should be done on a regular
basis to ensure that the sound
outlet does not become blocked.
If the HF3 wax fi lter becomes blocked, it will need to
be replaced.
• For HP receiver tubes, contact your hearing health-
care professional.
• For NP/LP receiver tubes, contact your hearing
healthcare professional or use the following instruc-
tions if you have been instructed to change the HF3
wax fi lter on the NP/LP receiver tube yourself.
Changing the HF3 wax fi lter (NP/LP receiver tube)
To remove the used wax fi lter:
1. Twist the metal end of the blue HF3 tool into the
wax fi lter on the hearing instrument approxi-
mately one and a half turns.
2. Pull the HF3 tool out and away from the receiver
tube. The wax fi lter should pull out with the tool.
If not, repeat Step #1, with an additional twist of
the tool.
21
21
43
1a
2
1b