BiliChek Service Manual
28
1014988
11.4 Hand Held Unit (HHU) Error Code List with Explainations
NOTE: All sub errors are displayed in the date field unless specified otherwise in each section. See Illustration at the top of
page 32.
Error: E01 Calibration
Sub Errors Description
000001 Reference component out of range. Install a fully charged battery pack and try several calibrations with
new BiliCal tips. If error persists, HHU requires service.
000002 Dynamic integration time out of range. Install a fully charged battery pack and try several calibrations
with new BiliCal tips. If error persists install a new measurement light bulb and try calibration several
times. If the error still persists, the HHU requires service.
000004 Shape index above threshold. Try a fully charged battery pack and a new BiliCal tip. If error persists,
HHU requires service.
000008 Peak pixel not found. Install a fully charged battery pack and try several calibrations with new BiliCal
tips. If error persists, HHU requires service.
008192 Stray Light Slope Error. Install a fully charged battery pack and try several calibrations with new BiliCal
tips. If error persists, HHU requires service.
016384 Mean of dark scan data above threshold. Install a fully charged battery pack and try several calibrations
with new BiliCal tips. If error persists, HHU requires service.
NOTE: Firware
Version 2.4/3.8
Only