Respironics bilichek - Service Manual

Respironics bilichek
36 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
Loading...
Service Manual
www.respironics.com

Other manuals for Respironics bilichek

Questions and Answers

  • P
    Paula PerryAug 17, 2025
    What to do if Respironics Measuring Instruments shows E03 error?
    • M
      mmartinAug 17, 2025
      If the Respironics Measuring Instruments displays 'E03' error, it could be due to the Disposable Tip being removed between calibration and measurement, or excessive ambient light. First, press and release F1 and F2 simultaneously. Then, remove the BiliCal, replace the tip, and recalibrate. If the error persists despite not removing the tip, the ambient light may be too high; move to a lower light environment and press F1 and F2 buttons together.
  • B
    Brandon DayAug 19, 2025
    How to fix low battery pack on Respironics Measuring Instruments?
    • J
      Julie VegaAug 19, 2025
      If your Respironics Measuring Instruments displays 'Low battery pack', replace the current battery with a fully charged one from the charger base. Then, place the discharged battery pack into well “B” for charging.
  • J
    James RamosAug 22, 2025
    What to do with discharged battery pack in Respironics Measuring Instruments?
    • A
      Anthony HarrisAug 22, 2025
      To address a 'Discharged battery pack' in your Respironics Measuring Instruments, replace the current battery with a fully charged battery from the charger base. Put the discharged battery pack into well “B” for charging.
  • R
    Reginald HarrisAug 25, 2025
    How to troubleshoot E01 error on Respironics Measuring Instruments?
    • M
      Mary MooreAug 26, 2025
      If the Respironics Measuring Instruments displays 'E01' error, press and release F1 and F2 at the same time. Make sure the BiliCal™ Tip is correctly placed. Press the TRIGGER button to restart calibration. If the error continues, try another BiliCal Tip. If it still persists, remove the BiliCal and clean the optical measurement tip (under the disposable tip) using an approved cleaning agent, then repeat calibration. If the error continues after cleaning, replace the measurement light bulb and recalibrate. If the 'E01' error persists after these steps, call for service.
  • G
    griffithpeterAug 27, 2025
    Why is battery pack not sensed in Respironics Measuring Instruments?
    • D
      dcolemanAug 27, 2025
      If your Respironics Measuring Instruments indicates 'Battery pack not sensed in well or charger base not connected to power supply', place the battery pack in the well and check all power connections to ensure they are properly connected.
  • Y
    ytapiaAug 30, 2025
    How to fix E00 error on Respironics Measuring Instruments?
    • M
      Megan SmithAug 30, 2025
      If the Respironics Measuring Instruments displays 'E00' error, install a charged battery pack into the battery compartment. Then, press F1 and F2 simultaneously, hold for one second, and release. This action will place the unit in the Set-up mode. Next, enter the settings specific to your institution.
  • D
    Daniel WillisSep 2, 2025
    What to do if Respironics Measuring Instruments displays E02 error?
    • S
      stephenhollandSep 2, 2025
      If the Respironics Measuring Instruments shows 'E02' error, press and release F1 and F2 at the same time, and then retest the patient. If the 'E02' error occurs again, replace the BiliCal, repeat the calibration, and retest the patient. If the error continues, call for service.
  • D
    David MurphySep 4, 2025
    What to do if Respironics Measuring Instruments shows E05 HIGH error?
    • S
      steven45Sep 5, 2025
      If the Respironics Measuring Instruments displays 'E05 HIGH', obtain a Total Serum Bilirubin level (TSB).
  • A
    Anthony EdwardsSep 7, 2025
    What does battery pack fully charged mean on Respironics Measuring Instruments?
    • C
      Christina MooreSep 7, 2025
      If the Respironics Measuring Instruments indicates 'Battery pack fully charged', it means the battery pack is ready for use.
  • P
    Peter PerezSep 10, 2025
    How to resolve E10 error on Respironics Measuring Instruments?
    • S
      Sarah WilliamsSep 10, 2025
      If the Respironics Measuring Instruments displays 'E10' error, replace the measurement light bulb. If the error continues after replacing the bulb, call for service.

Summary

Limited Warranty

Purpose of the Device

Theory of Operation

Warnings and Precautions

Description of the BiliChek System

System Specifications

Functional Description

Initial Set-Up

Set-Up Mode

Operation Of The Device

Performing a Calibration

Process for calibrating the BiliChek unit before each patient measurement.

Performing a Measurement

Steps to take patient measurements after calibration.

Troubleshooting Guide

Troubleshooting Guide

Lists error codes and their explanations for troubleshooting.

Troubleshooting Flow Chart

Visual guide to diagnose and resolve issues with the BiliChek system.

Drawings and Schematics

Contains electrical schematics and component layouts for the charger base.

HHU Error Code List with Explanations

Detailed explanations for error codes displayed on the Hand-Held Unit.

Routine (Operator) Cleaning

Maintenance and Repair and Replacement

HHU Measurement Light Bulb Replacement

Procedure for replacing the measurement light bulb in the Hand-Held Unit.

Charger Base Repair and Replacement

Steps for diagnosing, opening, and repairing or replacing the charger base.

Hand-Held Unit (HHU) Repair

Information on HHU repair, noting it's not field-serviceable except for the bulb.

Power Supply Repair;Replacement

Guidance on replacing the power supply unit when it fails.

Battery Pack Repair;Replacement

Instructions for battery pack replacement, as it is not repairable.

Software Upgrades

Respironics Customer Service and Product Support Contact Information