Warning Cautions
22 | P a g e
Read the entire manual before using the RespiWin 520
Install and configure the RespiWin 520 in accordance with the instructions provided in this
guide.
Do not calibrate the ventilator when it is connected to the patient, instead calibrate using
test lung given along with the machine.
Ventilator-dependent patients should be continuously monitored by qualified personnel or
adequately trained caretakers. These personnel and caretakers must be capable of taking
the necessary corrective action in the event of a ventilator alarm or malfunction.
Always connect OXYGEN (O2)After putting Ventilator on standby mode
Repairs and servicing of the device should only be performed by an authorized service
Representative.
Do not expose the device to excessive force, dropping or shaking.