Installation
8 | P a g e
1.Lets begin, while connecting 2 Connectors to the Inhale and Exhale Ports given on the left side
of the machine. Connector with the hole is the Expiratory Connector and without hole is the
Inspiratory Connector, so connect accordingly.
2. Now join 2 Valves with the Connector.
3. Next fix the Valves attached to the Patient Circuit along with the Flow Sensor.
4. Make sure HME Filter is fixed with the Flow Sensor
5.Near y piece of the Patient Circuit, there are 2 pipes (transparent and blue color) attached to the
Flow Sensor. Whichever color pipe is facing towards patient should be connected to “Flow Input
Negative” and second pipe should be connected to Flow Input Positive”.
6.Now we have to connect 2 Control Valve Pipes, Inhale pipe will be connected to the Inspiratory
Valve and Exhale Pipe will connected to the Expiratory Valve which is located on the side of the
machine
7. Finally, we will connect the power supply to the DC Input and Switch on the Unit.