9.0 Troubleshooting
*Contact Restek or your Restek representative for return instructions for servicing a damaged unit. Additional charges may
apply if the warranty has expired or the unit is damaged due to misuse.
SSuuggggeesstteedd SSoolluuttiioonnPPoossssiibbllee CCaauussee
PPrroobblleemm
Clean the probe tip to remove any debris
Visually inspect probe line for holes*
Recharge or return to Restek for battery replacement*
Probe clogged
Probe line punctured
Weak battery
Sensitivity decreased
Re-zero detectorDetector not zeroedResponse decreased
LED bar graph stays lit during operation
Detector re-zeroed before unit was purged out
Reference gas inlet covered by hand or
other object
Allow adequate time for detector to purge, then
re-zero
Remove obstruction