3. Technical Specifications
Environment
CAUTIONS
The equipment should be stored in draughty room free from corrosive gas or high
magnetic fields. If transport and storage environmental conditions go beyond the
required standard, the equipment must be placed in the operation environment for at
least 24 hours before starting operation.
Electric Specifications:
Supply VoltageAC100-240V,50/60Hz,90VA
Type of Protection
Against Electric ShockClass II Equipment
Degree of Protection
Against Electric ShockType BF Applied Part
Degrees of Protection
Provided by EnclosuresIP22
Mode of Operation Continuous
Physics Specification
Air OutletThe 22 mm conical air outlet complies with ISO 5356-1
DimensionsL× W × H) 205×154×197mm
Weight About 1.7kg
Humidifier
Water Capacity350ml, at the scale mark of maximum
Heater Setting1 to 6 (heating plate 30-75)
Pressure Drop<0.5 hPa at 60 LPM flow
Maximum Air Temperature<41
Humidity Output1040 mgH2O /L
Standard Compliance
IEC60601-1; IEC60601-1-2; ISO17510-1; ISO8185
Operation
Transport and Storage
Temperature
+5℃~ +35
-20℃~ +60
Humidity
10% 95% Non-condensing
10% 95% Non-condensing
Atmospheric
pressure
700hPa1060 hPa.
700 hPa1060 hPa.