User Manual_ResWell_CPAP
6
3. Technical Specifications
Environment:
CAUTIONS
:
The equipment should be stored in draughty room free from corrosive gas or high
magnetic fields. If transport and storage environmental conditions go beyond the
required standard, the equipment must be placed in the operation environment for at
least 24 hours before starting operation.
Electric Specifications:
Supply Voltage:AC100-240V,50/60Hz,≤90VA
Type of Protection
Against Electric Shock:Class II Equipment
Degree of Protection
Against Electric Shock:Type BF Applied Part
Degrees of Protection
Provided by Enclosures:IP22
Mode of Operation: Continuous
Physics Specification:
Air Outlet:The 22 mm conical air outlet complies with ISO 5356-1
Dimensions(L× W × H): 205×154×197mm
Weight: About 1.7kg
Humidifier:
Water Capacity:350ml, at the scale mark of maximum
Heater Setting:1 to 6 (heating plate 30-75℃)
Pressure Drop:<0.5 hPa at 60 LPM flow
Maximum Air Temperature:<41℃
Humidity Output:10~40 mgH2O /L
Standard Compliance:
IEC60601-1; IEC60601-1-2; ISO17510-1; ISO8185
10% ~ 95%( Non-condensing)
10%~ 95%( Non-condensing)