Pressure, Flow:
Pressure Range:4~20 cmH2O(in 0.5cmH2O increments)
Maximum Single Fault
Steady State Pressure:30 cmH2O
Pressure Tolerance: ±0.5 cmH2O
Sound Pressure Level: <30dB (A) when the device is working at the pressure of 10cmH2O
Test pressures
4
8
12
16
20
Measured pressure at the patient
connection port(hPa)
3
7
11
15
19
Average flow at the patient connection
port(l/min)
36.8
41.3
49
45.5
55.2