Chapter 2 Performance Verification Test RIGOL
DG4000 Performance Verification Guide
Chapter 2 Performance Verification Test
Taking DG4162 as an example, this chapter introduces the method of performance
verification test for DG4000 series function/arbitrary waveform generator.
Test items include:
Frequency Accuracy Test
AC Amplitude Accuracy Test
DC Offset Accuracy Test
AC Flatness Test
Harmonic Distortion Test
Rise/Fall Time Test
Spurious Signal Test
Overshoot Test
Note:
1) Please make sure the instrument has been warmed up for at least 30 minutes
under the specified operational temperature (18℃ to 28℃) before the test.
2) Any of the following tests must be done under the specified operational
temperature (18℃ to 28℃).
3) Please reset the instrument to the factory setting before or after executing any
of the following tests.