RIGOL  Chapter 2 Front Panel Operations 
2-60     DG800 User's Guide 
Ramp: 1 μHz to 1 MHz 
Arb: 1 μHz to 10 MHz 
  The range of the frequency span is affected by the center frequency. 
Center frequency < 
: the range of the frequency span is ±2× (center 
frequency – 
); 
Center frequency ≥ 
: the range of the frequency span is ±2× (
 - 
center frequency). 
 
Take sine as an example. 
 is 1 μHz, 
 is 35 MHz, and 
 is 17.5 MHz. 
If the center frequency is 550 Hz, the frequency span range is ±2×(550 Hz–1 μHz) = 
±1.099999998 kHz; if the center frequency is 20 MHz, the frequency span range is 
±2×(35 MHz – 20 MHz) = ±30 MHz. 
 
The generator will restart to sweep from the specified "start frequency" after the 
"center frequency" or "frequency span" is modified. 
 
 
In large-scale sweep, the amplitude characteristic of the output signal might 
 
 
Sweep Trigger Source 
 
The trigger sources of the sweep can be Internal, External, or Manual. When the 
signal generator receives a trigger signal, a sweep output is generated, and then the 
signal generator waits for another trigger signal. After the sweep function is enabled, 
tap the Source to select "Internal", "External", or "Manual". The default is 
"Internal". 
 
1.  Internal Trigger 
The generator outputs continuous sweep waveforms when the internal trigger is 
selected. The trigger period is determined by the specified sweep time, return 
time, start hold, and stop hold time. 
 
2.  External Trigger 
In the External trigger, the signal generator receives the trigger signal input from 
the rear-panel [CH1/Sync/Ext Mod/Trig/FSK] connector. Once it receives a 
TTL pulse with a specified polarity, a sweep is initiated. To specify the polarity of 
the TTL pulse, tap the Trig In to select "RisEdge" or "FallEdge". By default, it is 
"RisEdge".