Chapter 1 Test Overview RIGOL
DL3000 Performance Verification Manual
5. Keep the cables of the test probe as short as possible.
CAUTION
The tests described in this manual are required to be carried out by the
professionals. During the test, dangerous voltages might be present on
the output terminals of the power supply and some test devices.
Recommended Test Devices
You are recommended to use the devices listed in Table 1-2 or other devices whose
performance specifications meet the "Performance Requirement" listed in Table 1-2.
Table 1-2 Recommended Test Devices
Device
Performance
Requirement
Recommended
Model
Usage
Digital
Multimeter
Readout Resolution: 6½
RIGOL DM3068
Measures DC
voltage
DC Power Supply
DC 150 V/1 A
DC 8 V/90 A
DSP-150-005HD
DSP-008-090HD
Provides voltage
and current
Current
Sampling
Resistor R
M
0.01 Ω (temperature drift: 1
ppm)
RUG-Z-R010-01-TK1
Measures
programming
accuracy and
readback accuracy
USB Cable
--
--
Realizes remote
control, measures
programming
accuracy and
readback accuracy

Other manuals for Rigol DL3000 Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Rigol DL3000 Series and is the answer not in the manual?

Rigol DL3000 Series Specifications

General IconGeneral
BrandRigol
ModelDL3000 Series
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals