Chapter 2 Specifications Test RIGOL
DL3000 Performance Verification Manual
2-25
Test Record Forms:
Table 2-9 Test Record Form of DL3021 CR Programming Accuracy
Model
DL3021 (150 V/40 A)
The resistance range of the electronic load is from 0.08 Ω to 15 Ω.
Specification
Programming Accuracy
Vin/Rset*(0.2%)+0.08 A
DL3021 Setting
Resistance
R
L
[1]
Test Result
U
1
U
2
R
1
=U
1
/(U
2
/R
M
[2]
)
Programming Accuracy: |R
1
- R
L
[1]
|
DL3021 Setting
Resistance
R
H
[3]
Test Result
U
3
U
4
R
2
=U
3
/(U
4
/R
M
[2]
)
Programming Accuracy: |R
3
- R
H
[3]
|
The resistance range of the electronic load is from 2 Ω to 15 kΩ.
Specification
Programming Accuracy
Vin/Rset*(0.2%)+0.08 A
DL3021 Setting
Resistance
R
L
[1]
Test Result
U
1
U
2
R
1
=U
1
/(U
2
/R
M
[2]
)
Programming Accuracy: |R
1
- R
L
[1]
|
DL3021 Setting
Resistance
R
H
[3]
Test Result
U
3
U
4
R
2
=U
3
/(U
4
/R
M
[2]
)
Programming Accuracy: |R
3
- R
H
[3]
|
Note
[1]
: indicates the set low resistance value.
Note
[2]
: R
M
= 0.01 Ω.
Note
[3]
: indicates the set high resistance value.

Other manuals for Rigol DL3000 Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Rigol DL3000 Series and is the answer not in the manual?

Rigol DL3000 Series Specifications

General IconGeneral
BrandRigol
ModelDL3000 Series
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals