RIGOL Appendix
DL3000 Performance Verification Manual
1 (optional, USB-GPIB interface converter)
5% to 80% RH (without condensation)
Certification Information
Complies with the requirements of the following directive and
standards:
EMC Directive 2014/30/EU
EN 61326-1
EN 61000-3-2
EN 61000-3-3
Under following condition:
The maximum length of all connecting cables and wires to the
DL3000 series are less than 3 m.
Complies with the requirements of the following directive and
standards:
Low Voltage Directive 2014/35/EU
EN 61010-1: 2010
IP Degree: IP20
Pollution Degree: PD 2
Over Voltage Category: OVC II
Operation Location: Indoor use only; not for wet condition
Note
[1]
: Data measured after 30-second current sinking at the programming value (applicable to
the programming accuracy in CC mode and CV mode).
Note
[2]
: FS indicates the full scale.
Note
[3]
: The input voltage for the low range in CR mode should be smaller than 8 V.
Note
[4]
: The programming accuracy in CR mode is also determined by the input voltage accuracy.
Note
[5]
: Current slew rate: rising slew rate for 10%~90% of the current (0-maximum current).
Note
[6]
: When the input voltage is greater than 5 V, the maximum current slew rate is 5 A/μs;
when the input voltage is greater than 4 V, the maximum current slew rate is 2.5 A/μs; when
the input voltage is greater than 2 V, the maximum current slew rate is 0.1 A/μs.
Note
[7]
: Following a steady 30-minute current sinking, change in current/voltage sinking over 8
hours under constant load, line, and ambient temperature.