RIGOL Appendix
DL3000 Performance Verification Manual
14
CP Mode
Model
Range
Resolution
DL3021
0 ~ 200 W
100 mW
DL3021A
DL3031
0 ~ 350 W
100 mW
DL3031A
Con Mode
Model
Frequency Range
Frequency
Resolution
Frequency
Accuracy
Duty Cycle Range
DL3021
0.001 Hz~15 kHz
0.8%
± 0.5%
5%~95%, 1%
DL3021A
0.001 Hz~30 kHz
DL3031
0.001 Hz~15 kHz
DL3031A
0.001 Hz~30 kHz
Current Slew Rate
[5]
Model
Range
Resolution
Accuracy
DL3021
0.001 A/μs~0.25 A/μs
0.001 A/μs~2.5 A/μs (>5 V)
[6]
0.001 A/μs
5%+10 μs
DL3021A
0.001 A/μs~0.3 A/μs
0.001 A/μs~3 A/μs (>5 V)
DL3031
0.001 A/μs~0.25 A/μs
0.001 A/μs~2.5 A/μs (>5 V)
0.001 A/μs
5%+10 μs
DL3031A
0.001 A/μs~0.5 A/μs
0.001 A/μs~5 A/μs (>5 V)
Current Readback
Model
Range
Resolution
Accuracy
Temperature Coefficient
DL3021
0 ~ 40 A
1 mA
± (0.05%+0.05%FS)
50 ppm/
DL3021A
0.1 mA
DL3031
0 ~ 60 A
1 mA
± (0.05%+0.05%FS)
50 ppm/
DL3031A
0.1 mA
Voltage Readback
Model
Range
Resolution
Accuracy
Temperature Coefficient
DL3021
0 ~ 150 V
0.1 mV
± (0.05%+0.02%FS)
20 ppm/
DL3021A
DL3031
0 ~ 150 V
0.1 mV
± (0.05%+0.02%FS)
20 ppm/
DL3031A